핀 포토다이오드 라돈 검출기의 온습도 보정에 관한 연구
출처
대한전자공학회
발행일
2020-06-30
분야분류
IT소재부품, 해외동향, IT부품정보, IT부품정보 네트워크,
원문정보
원문 바로가기
The radon concentration values measured by the PIN photodiode radon detectors in indoor is influenced by temperature and humidity, but few studies have focused on the calibration method for effects of temperature and humidity on the measured values using PIN photodiode radon detector. In this paper, temperature, relative humidity, radon count, and radon concentration were measured under different conditions, and we derived the multiple linear regression model for calculation of the radon concentration using the measurement results and multiple regression analysis. We compared the RMSE values for calibrated case and noncalibrated case for PIN photodiode radon detector and found that the RMSE value of calibrated case was smaller than that of noncalibrated case.